Accidental fire reported in chest cavity of patient during emergency cardiac surgery-case report

Published On 2019-07-24 12:30 GMT   |   Update On 2019-07-24 12:30 GMT

Dr Ruth Shaylor and colleagues from Austin Health have reported a case wherein while the doctors were doing open-heart surgery on the person, his chest cavity suddenly and inexplicably bursts into flames. They presented the case at Euroanaesthesia Congress in Vienna as the case of a 60-year-old man who suffered a flash fire in his chest cavity during emergency heart surgery.


Dr Ruth Shaylor and colleagues from Austin Health in Melbourne, Australia, where the incident took place, warn that the case highlights the potential dangers of dry surgical packs in the oxygen-enrich environment of the operating theatre where electrocautery devices (using heat to stop vessels from bleeding) are used.


In August 2018, a 60-year-old man presented for emergency repair of an ascending aortic dissection--a tear in the inner layer of the aorta wall in the chest. The patient had a history of chronic obstructive pulmonary disease (COPD) and had undergone coronary artery bypass grafting one year previously.


As surgeons began to operate, they noted that the man's right lung was stuck to the overlying sternum with areas of overinflated and destroyed lung (bullae; often caused by COPD). Despite careful dissection, one of these bullae was punctured causing a substantial air leak. To prevent respiratory distress, the flows of anaesthetic gases were increased to 10 litres per minute and the proportion of oxygen to 100%.


Soon after, a spark from the electrocautery device ignited a dry surgical pack. The fire was immediately extinguished without any injury to the patient. The rest of the operation proceeded uneventfully and the repair was a success.

"While there are only a few documented cases of chest cavity fires--three involving thoracic surgery and three involving coronary bypass grafting--all have involved the presence of dry surgical packs, electrocautery, increased inspired oxygen concentrations, and patients with COPD or pre-existing lung disease", explains Dr Shaylor.


"This case highlights the continued need for fire training and prevention strategies and quick intervention to prevent injury whenever electrocautery is used in oxygen-enriched environments. In particular surgeons and anaesthetists need to be aware that fires can occur in the chest cavity if a lung is damaged or there is an air leak for any reason, and that patients with COPD are at increased risk."

Thankfully, the fire was immediately extinguished without any injury to the patient and the rest of the operation proceeded uneventfully and the repair was a success.

Tags:    

Disclaimer: This site is primarily intended for healthcare professionals. Any content/information on this website does not replace the advice of medical and/or health professionals and should not be construed as medical/diagnostic advice/endorsement or prescription. Use of this site is subject to our terms of use, privacy policy, advertisement policy. © 2020 Minerva Medical Treatment Pvt Ltd